Technical Data
Microbiological analyser-
Micro- organism detection based on impedance splitting method.
Dry - Incubator
Each BacTrac 4300 unit contains two 32 sample well, dryblock incubators
that are independently thermostatted.
Up to 12 units ( 768 wells, 24 incubators ) can be controlled by
one PC. Each sample well has high reliability electrical contacts
that accommodate either disposable or reusable measurement cells.
Each incubator section is covered by a convenient , easily cleaned
glass cover.
Temperature Control
The temperature range of each 32 well incubator section can be adjusted
in software by the control PC from 0º C to +65ºC and maintained
with an accuracy of +/- 0.1 ºC.
In addition to temperature information being available on the control
PC, each BacTrac 4300 unit has a 2 line vacuum fluorescent display
that provides cell location and temperature information. For incubator
temperatures from 0ºC to 7º C above room temperature, an optional
water circulation cooler is required.
Measuring Cells
Direct impedance measurements for bacteria can be made using either
20 ml disposable plastic or 10 ml reusable glass cells. Glass cells
and their electrode assemblies are autoclavable. Disposable plastic
cells are sterile and pre-assembled. Custom formulated impedance
media is available for reconstitution and filling by the user or
cells can be ordered pre-filled. Indirect impedance measurements of yeasts and moulds can be made using nested cells: a 20 ml reusable outer cell with integral electrodes used with a disposable plastic (7ml) inner vial that mates with it.
Dimensions, weight, power consumption
Dimensions: 430 mm (W) x 640mm (D) x 380mm (H) (16.9 x 25.2 x 15.0
inches). Weight: 34 Kg (75 lbs.). Power requirements: 230 V, 3.9
A, 50 Hz or 115 V, 7.8 A, 60 Hz. (Please specify with order). Dimensions,
weight and power requirements are for the BacTrac 4300 unit only
and do not include the control PC.
Electronics
Each BacTrac 4300 has a built-in computer subsystem for incubator
temperature control, impedance measurements, fault monitoring and
bi-directional communication with the control PC. Impedance measurements
are made at 1 KHz and each sample cell location is polled every
10 minutes.
Incubator temperature set points are controlled by the PC and maintained
by the 4300's on board computer.
Minimum Computer Requirements
Intel™ Pentium IV or comparable processor from another manufacturer, min 2.2 GHz, 1024 MB RAM, VGA >16 MB, HD >10 MB, CD-ROM, serial interface card necessary (available from SY-LAB) if more than one BacTrac connected to PC, monitor: min 17" (1-6 incubators), 19" (recommended for use with 7-12 incubators), min resolution 1024 x 768 pixels, colour printer, Windows XP Professional operating system..
Software
MS Windows®, menu driven, automatic time-date registration, continuous sample monitoring, colour-coded detection time thresholds and evaluation status indication, rapid identification of analysis status.
Allocation (including for individual positions) of analysis parameters and calibration curves, easy text input and editing, flexible indication and sorting parameters, report production, automatic periodical backup via service program, LAN connection, data export opportunities, data interfaces to external programs available.
|